TM 5-4220-227-14
MEDICAL LOCK (15). Provides access to I/L for small sustenance items and medical supplies.
OUTER LOCK (16). Allows personnel access to primary treatment area.
INNER LOCK (17). Contains primary pressurized treatment area.
1-9. Equipment Data.
a. Recompression Chamber.
Manufacturer...............................................................................CVI Inc.
Model..........................................................................................RC-100-2*
Part Number ...............................................................................C-867-7300-81*
Weight ........................................................................................3000 lbs
Length.........................................................................................145 in.
Width ..........................................................................................60 in.
Height .........................................................................................76 in.
b. Heater/Chiller Module.
Manufacturer...............................................................................Amron International
Model..........................................................................................9000 ECS
Weight ........................................................................................175 lbs (79.38 Kg)
Length.........................................................................................19.5 in. (495.3 mm)
Width ..........................................................................................27.0 in. (685.8 mm)
Height .........................................................................................30.25 in. (768.35 mm)
* As modified by Cage Code (80064) Drawing No. 330014.
c. Electrical Power Source Requirements. An electrical power source rated at 110 VAC, 60 Hertz, 20 Amps, is
required to power the 3-person recompression chamber. This source is connected to the Ground Fault Interrupter (GFI).
Two additional electrical power sources are required to power the heater/chiller module located left rear of, and behind
the chamber. A 220 VAC, 60 Hertz, 20 Amp source is required to power the 3 HP electrical motor for the compressor,
and a 110 VAC, 60 Hertz, 10 Amp source is required to power the 1/3 HP electrical motor for the fluid circulation pump.
1-7