TM 5-4220-227-14
Component
Index
Step
(Location)
No.
Procedure
Check
Note
20.
OXH-030 through OXH-032
12
Shut
(02 Bank)
through
14
21.
OXH-033 (02 Bank #2)
16
Open
22.
OXH-034 (02 Bank #2)
17
Open
23.
OXH-035 (02 Bank #2)
18
Open
24.
OGA-28 (02 Bank)
19
Record bank #2 pressure from
OGA-028 here
25.
OXH-033, OXH-034,
16
Leave open if bank #2 is lower than
OXH-035
through
bank #1, if not, shut
18
26.
OXH-030, OXH-031,
12
Leave shut unless bank #1 is lower
OXH-032
through
pressure than bank #2
14
27.
Oxygen Supply
NA
Bank #
is on line with PSIG
28.
OHL-024 (Console)
20
Dial in until OGA-025 reads
80-100 PSIG
29.
OLP-020 (Console)
21
Open
30.
OLP-021 (Console)
22
Open
31.
OEX-037 (Console)
23
Open
32.
OEX-038 (Console)
24
Open
33.
Ground Fault Interrupter
25
Plug into power
34.
SWT-020 (Console)
21
Turn on - push SWT-021 to test.
SWT-020 should turn off automatically.
Turn SWT-020 back on.
35.
SWT-019 (Console)
26
Turn on
36.
SWT-001 (Console)
27
Turn on
Figure 2-10. System Startup Checklist (Sheet 2 of 11).
2-23